| Variation ID | Pack Size | List Price | Qty | ||
|---|---|---|---|---|---|
| photo_camera | 1 Pack
L300865 |
£6313.90 |
For clinical laboratory analyser feeder
Type II / Deionised (DI) water
Designed for use in laboratory applications, such as HPLC, ion chromatography, atomic absorption, hydrogen generation and clinical analyser feed, the Purite HP produces a consistent flow of purified water to a quality better than 10MΩ.cm.
GP units are not supplied with a tank, they are designed to be used in conjunction with external 50 or 100L tanks - available here
Output
Dispense rate of up to 2 litres/min.
Quality of Water Produced
Type II deionised water. Guaranteed > 10MΩ.cm water quality.
Suitable Applications
General Lab use such as glassware rinsing, buffers and stains, reagent make-up, and media preparation.
Bench or Wall Mounting Ability
Purite units can easily be wall-mounted and are simply connected by a conventional hose to a laboratory tap. They operate under standard water pressure, so no additional parts are required.
Clear Display
Easy to use LCD – Colour touch screen.
Energy Saving
Features an energy saving intelligent stand-by mode.
Unit Specification
| Specification | 160 | 320 |
|---|---|---|
| Width (mm) | 440 | 440 |
| Depth (mm) | 560 | 560 |
| Height (mm) | 750 | 750 |
| Max shipping weight (kg) | 36 | 41 |
| Max working weight (kg) | 51 | 59 |
| Requirement | Value |
|---|---|
| Power | Single Phase, 110–230V, +/-10%, 50/60 Hz |
| Feed water | Potable |
| Maximum TDS (ppm) | 1000 |
| Minimum inlet pressure – psi (bar) | 30 (2.1) |
| Maximum inlet pressure – psi (bar) | 90 (6.2) |
| Feed water temperature | 1–35°C |
| Temperature | 160 (l/hr) | 320 (l/hr) |
|---|---|---|
| @ 10°C | 14.4 | 30 |
| @ 25°C | 24 | 48 |
Product outputs based on a feed water pressure of 4 bar.
| Feature | Details |
|---|---|
| Pure water storage | 20 litre storage tank as standard (External 50 & 100 litre tanks available) |
| Display panel | LCD – Colour touch screen |
| Pre-treatment cartridge | ✓ |
| Reverse osmosis | ✓ |
| Deionisation cartridge | ✓ |
| Internal filtration | Microfiltration |
| Point of use | UV / LED |
| UV lamp | ✓ |
| Recirculation pump | ✓ |
| Ultrapure polishing cartridge | – (External tank version only) |
| Parameter | Value |
|---|---|
| Inorganics | >10 MΩ·cm |
| pH† | Neutral |
| Bacteria | <1 cfu/ml |
| Organics – TOC (ppb) | <20 |
| Particles | 0.005 μm |
| Endotoxins | – |
| DNases | – |
| RNases | – |
| Parameter | Value |
|---|---|
| Dispense modes | Latched – hold |
| Dispense flow rate | Up to 2.0 l/min |