Variation ID | Product Code | Pack Size | List Price | Qty | ||
---|---|---|---|---|---|---|
photo_camera | HI-98165 | 1 Pack
HI-98165 |
£873.10 |
The HI-98165 is a rugged, waterproof, Foodcare portable pH meter that measures pH and temperature using the FC2423 cheese pH electrode.
FC2423:
Cheese products can provide challenges for the person that needs to measure pH. Cheese products tend to be solid to semi-solids. Both types of samples will coat the sensitive glass membrane surface and/or clog the reference junction. The FC2423 supplied with the HI-98165 is designed specifically for measuring pH in cheese.
From a conic tip shape in a durable 5 mm diameter stainless steel body for easy penetration into cheese without leaving a large hole to an open junction that resist clogging; the FC2423 is an ideal general-purpose pH electrode for cheese. The FC2423 connects to the HI-98165 with a quick-connect, waterproof DIN connector, allowing for a secure, non-threaded attachment.
The robust stainless steel sheath paired with the conical sensing tip allows for penetration in cheese at various points throughout the production process. An integrated temperature sensor also ensures that all pH measurements are compensated for temperature without the need for a separate temperature probe.
The HI-98165 has the Hanna unique CAL Check feature that alerts the user to potential problems during the calibration process. This is a very important for the food processor since it is likely that the probe will be coated with the solids found in the food product being measured. This coating can easily lead to errors in pH measurement. By comparing previous calibration data to the current calibration, the meter will inform the user, with display prompts, when the probe needs to be cleaned, replaced, or if the pH buffer might be contaminated. After calibration, the overall probe condition is displayed on screen as a percentage from 0 to 100% in increments of 10%. The probe condition is affected by both the offset and slope characteristics of the pH electrode, both of which can be found in the GLP data.
AutoHold
Pressing the AutoHold virtual key in measurement mode, the meter will freeze and automatically log a stable reading. An “out of calibration range” warning can be enabled that will alert the user when a reading is not within the bracket of calibrated pH values.
Help menu
A contextual help menu based on the screen that is currently being viewed can be accessed at any time by the press of a dedicated button.
Log-on demand
The log-on-demand mode allows the user to record and save up to 200 samples. The logged data, along with the associated GLP data, can then be recalled or transferred to a PC with Hanna’s HI-920015 micro USB cable and HI-92000 software for traceability in record keeping for specific product batches. GLP data includes date, time, calibration buffers, offset, and slope, and is directly accessible by pressing the dedicated GLP key.
LCD screen
The high contrast, graphic LCD screen is easy to view outdoors in bright sunlight as well as in low-lit areas with the backlight. A combination of dedicated and virtual keys allows for easy, intuitive meter operation in a choice of languages.
pH range | - | -2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000pH | ||||
pH Resolution | - | 0.1 pH; 0.01 pH; 0.001 pH | ||||
pH Accuracy | - | ±0.1; ±0.002 pH | ||||
pH Calibration | - | up to five point calibration, seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) + five custom buffers | ||||
mV range | - | ±2000mV | ||||
mV Resolution | - | 0.1 mV | ||||
mV Accuracy | - | ±0.2 mV | ||||
Temperature Range | - | -20.0 to 120.0˚C | ||||
Temperature Resolution | - | 0.1˚C | ||||
Temperature Accuracy | - | ±0.4˚C (excluding probe error) | ||||
Temperature Compensation | - | manual or automatic from -20.0 to 120.0˚C | ||||
Slope Calibration | - | from 80 to 110% | ||||
Electrode/Probe | - | FC2423 Foodcare Stainless Steel body, pH electrode with internal temperature sensor, quick connect DIN connector and 1m (3.3ft cable) | ||||
Input Impedance | - | 1012 Ohms | ||||
Log-on demand | - | 200 samples (100pH and 100mV range) | ||||
PC Connectivity | - | opto-isolated USB with optional HI-92000 software and micro USB cable | ||||
Battery Type/Life | - | 1.5V AA batteries (4) / approximately 200 hours of continuous use without backlight (50 hours with backlight) | ||||
Auto-off | - | user selectable: 5, 10, 30, 60 min or can be disabled | ||||
Environment | - | 0 to 50˚C; RH 100% (IP67) | ||||
Dimensions | - | 185 x 93 x 35.2mm | ||||
Weight | - | 400g |