Atom Scientific Ltd Logo webp
shopping_cart 0 search menu

HI-2630 Conductivity Benchtop Meter

HI-2630 Conductivity Benchtop Meter Group Picture
Code: HI-2630-02
  Pack Size List Price Qty  
photo_camera 1 Pack
HI-2630-02
£525.00
Pack: 1 Pack

£525.00

HI-2630 benchtop meter enables fast, accurate measurements utilising the Hanna Instruments digital electrode for conductivity (HI-763100) which includes an integral temperature sensor. Each digital electrode has a unique serial number and the parameter that it measures is automatically identified by the meter.

The intuitive design simplifies configuration, calibration, measurement, data logging and transfer of data to a USB thumb drive or computer.

The meter has a large display with capacitive keys which features a 5.5” (14 cm) LCD. The large display provides a 130 ° wide viewing angle.

Key Features:

  • Automatic parameter recognition
  • EC choice of measurement unit: Conductivity: µS/cm, mS/cm
  • TDS - ppm (mg/L), g/L
  • Adjustable resolution for EC measurements
  • Dedicated GLP key
  • Simplified data transfer to a PC
  • Internal clock and date

Digital probe input - simply connect each probe via the 3.5 mm jack, Digital Smart Electrodes are automatically recognised.

Clear, full text readout - The HI-2630 presents clear, detailed text prompts at the bottom section of the screen.

GLP - The sensor retains data from the most recent calibration, including date, time, and buffers utilised.

Capacitive touch keypad - HI-2630 features sensitive capacitive touch buttons for accurate keystrokes for navigating menus and screens. Since they are part of the screen, the buttons can never get clogged with sample residue.

Easy to read LCD - HI-2630 features a 5.5” (14 cm) LCD display that you can clearly view from over 5 m (16.4’). The large display, with its wide 130° viewing angle, provides one of the easiest to read LCDs in the industry.

Meter is supplied complete with:

  • HI-763100 digital conductivity electrode
  • HI-764026 Electrode holder with the following accessories: base plate with integrated pivot pin; cable holder clip, attached
  • 12880 µS/cm calibration solution, x 2 sachets
  • 1413 µS/cm calibration solution, x 4 sachets
  • 5000 µS/cm calibration solution, x 2 sachets
  • Electrode rinse solution x 2 sachets
  • HI-920018 - 1.5m USB-C to USB-C cable
  • USB-C power adapter
  • Instrument quality certificate
  • Probe / electrode quality certificate
  • Battery safety insert
  • Quick reference guide with QR code for instruction manual download

Warranty - Meter has 2 years. 6 months from date of purchase for electrodes and probes

probe  
   EC Range 0.00 to 29.99 µS/cm;
30.0 to 299.9 µS/cm;
300 to 2999 µS/cm;
3.00 to 29.99 mS/cm;
30.0 to 200.0 mS/cm;
up to 500.0 mS/cm,
and absolute conductivity (TDS) is the conductivity value without temperature compensation
 Resolution 0.01 µS/cm;
0.1 µS/cm;
1 µS/cm;
0.01 mS/cm;
0.1 mS/cm
 Accuracy (@25°C) ±1% of reading (±0.5 µS/cm or 1 digit, whichever is greater)
 Calibration Cell factor calibration; six standards available: 84µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 118.8 mS/cm, one point offset: 0.00 µS/cm
   Temperature Compensation Automatic, -5 to 100.0°C; NoTC, can be selected to measure absolute conductivity
 Temperature Coefficient 0.00 to 6.00%/°C (for EC and TDS only), default value is 1.90%/°C
TDS - using HI-763100 four-ring conductivity probe  
   TDS Range 0.00 to 14.99 mg/L (ppm);
15.0 to 149.9 mg/L (ppm);
150 to 1499 mg/L (ppm);
1.50 to 14.99 g/L;
15.0 to 100.0 g/L;
up to 400.0 g/L absolute TDS using 0.80 conversion factor
 Resolution 0.01 mg/L (ppm);
0.1 mg/L (ppm);
1 (ppm);
0.01 g/L;
0.1 g/L
 Accuracy (@25°C) ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
 Calibration through EC calibration
 TDS Factor in Standard Mode 0.40 to 0.80 (default value is 0.50)
Temperature  
   Temperature Range -20.0 to 120.0 °C
 Resolution 0.1°C
 Accuracy ±0.5°C
Additional Specifications  
PC Interface USB-C
Storage Interface USB-C
Power supply USB Type C (5 VDC; 500mA)
Environment 0 to 50˚C; Maximum 95% RH non-condensing
Dimensions 205 x 160 x 77 mm
Weight Approximately 0.85 kg

You may also like...

Cleaning solution pH probes, Hanna

GL890 Blue Nitrile Gloves, Powder Free, XS-XXL

Conductivity standard solution 84uS/cm

Conductivity standard solution 1413uS/cm Hanna

Atom Scientific Ltd